What is it about?

Analytical interferences in immunoassays keep being a problem on routine analysers. We show a new case of anti-streptavidin interference with the measurement of free T4 and free T3 on Cobas 8000 FT4 II and FT3 III assays. The interference diminished with the new FT4 III assay.

Featured Image

Read the Original

This page is a summary of: Alarmed by misleading interference in free T3 and free T4 assays: a new case of anti-streptavidin antibodies, Clinical Chemistry and Laboratory Medicine (CCLM), October 2019, De Gruyter,
DOI: 10.1515/cclm-2019-0845.
You can read the full text:

Read

Contributors

Be the first to contribute to this page