This is a prospective cohort study of intensive care unit patients who received a comparative venous glucose measurement along with up to three capillary measurements. We found only one clinical significant measurement inaccuracy and found that hypoglycemia was never overestimated.

Featured Image

Why is it important?

While previous studies have focused on the numerical differences between measurement modalities, ours focused on whether those differences actually matter. We noted no significant clinical impact in our cohort and are unlikely to abandon the use of convenient and fast POC glucometers in our intensive care unit.

Read the Original

This page is a summary of: Clinical Reliability of Point-of-Care Glucose Testing in Critically Ill Patients, Journal of Diabetes Science and Technology, July 2019, SAGE Publications,