What is it about?

We measured the time required for sham nitro to reach the end of the cannula in common infusion devices: 18g PIVC, 15g 45mm IO, 7F Triple lumen CVC, 5F introducer catheter and the effect of a 30mL drive or carrier line on the time-to-infusion.

Featured Image

Why is it important?

Many of us do not know how dead space effects infusion time. There are many high concentration, low rate of administration drugs used in resuscitation and critical care. We performed this study to illustrate the relationship between time to drug, dead space aka priming volume and carrier lines.

Perspectives

Infusion technique is central to the role of nurses in resuscitation and critical care. Delays of second to minutes can have disastrous effects. Strategies to minimize delays are absent from infusion guidelines & manuals. The findings of our study are applicable to any high concentration, low rate of infusion medication such as vasopressors or vasodilators.

Mr Matthew John Douma
Royal Alexandra Hospital

Read the Original

This page is a summary of: How intravenous nitroglycerine transit time from bag-to-bloodstream can be affected by infusion technique: a simulation study, Emergency Medicine Journal, April 2015, BMJ,
DOI: 10.1136/emermed-2014-204523.
You can read the full text:

Read

Contributors

The following have contributed to this page